Description
The Protector Echo Filtered Fume Hood requires no ducting and is available in 4, 5, 6 and 8 ft widths. Neutrodine® Unisorb Filtration Technology offers comprehensive molecular filtration. eGuard® communication software offers real-time monitoring on the users phone or computer. From a remote location, data such as usage authorization, filter saturation detection, sash position and temperature may be observed and managed.
Erlab-proven Neutrodine® Unisorb Filters are the heart of GFH Filtration Technology. Models are available for carbon-only, carbon with acid sensor, carbon with formaldehyde sensor, or HEPA-only filtration. Filters are sold separately.
Features (All Models)
• Eco-Foil™ air foil with aerodynamic Clean-Sweep™ airflow openings
• Cord-Keeper™ Slots on left and right side of air foil
• Removable front and side panels, and front and interior service access panels for access to plumbing and electrical wiring
• By-pass airflow design
• Glacier white powder-coated steel exterior
• Chemical-resistant, fiberglass-reinforced, composite panel liner
• Pre-wired vapor-proof LED lights for up to 63 foot candles rated for 50,000 hours
• eGuard2 software communication
• Factory-installed command module, fan button and light button
• Roughing prefilter
• Built-in exhaust fans to maintain 60 fpm face velocity at 16 in. sash opening height
• Sound pressure of <60 dB(A) with sash at operating height and <48 dB(A) with sash closed
• Sensor package for primary filter breakthrough of solvent and acid fumes, laboratory air quality, sash position, and temperature
• Audible/visual alarms for breakthrough detection, temperature 40° C (104° F), fan failure and high sash position opening. At 60°C (140°F), all fans stop
• Single piece, tempered safety glass vertical-rising sash with cable pulley, sash weight and powder-coated aluminum sash handle and tracks with airflow openings. Sash stop at 16 in. sash opening height on 4 ft models and 13 in. sash opening height on 3 ft models. Sash opens to 26 in. high for loading
• Factory-prepared for up to two electrical duplex receptacles and four service fixtures
• 28.5 in. high sightline from the work surface and header panel
• Power cord with plug
Required Accessories (Not Included)
• Neutrodine® Unisorb Filters (except HEPA only models which require HEPA filters). Number of filters required is determined by the width of the hood
• Work Surface
• Base Cabinet or Stand
Options Available Include
• HEPA filter
• Region-specific electrical plug configurations
• Service fixtures
Compliance
• SEFA 9, DHIII
• SEFA 8, Cabinet Surface Finish Tests
• Modified ASHRAE 110
• ANSI Z9.5-2011
• AFNOR NF-X 15-211
• ASTM E84-09C
• UL Standard 61010-1
• CAN/CSA C22.2 No. 61010-1
Specifications
• Airflow Type: Bypass
• Product Subcategory: General Chemistry
• Air foil Type: Eco-Foil
• Blower Requirements: Integral blower included
• Conformance: AFNOR NF-X, ANSI Z9.5, ASTM E84, CAN/CSA C22.2, Modified ASHRAE 110, SEFA 9, UL 61010
• Enclosure Height: 54.2 in.
• Lighting: LED
• Power Cord & Plug: 11.75 ft power cord with plug
• Sash Type: Cable and pulley
• Service Fixtures: 0
• Style: Benchtop
Image(s) are representative of the product group and not necessarily the individual product.